User`s manual

Cerebus® User’s Manual
Revision 13.00 / LB-0028 Page 9
Table 206 Recommended separation distances between portable and mobile RF communications equipment
and the EQUIPMENT or SYSTEM - for EQUIPMENT or SYSTEMS that are not LIFE-SUPPORTING (refer to
60601-1-2).
Recommended separation distances between portable and mobile RF communications equipment and
the Cerebus® system
The Cerebus® system is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Cerebus® system can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the Cerebus® system as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
P
V
d ]
5,3
[
1
P
E
d ]
5,3
[
1
P
E
d ]
7
[
1