Specifications

Cerebus
TM
Stim Switch Users Manual
Rev. 1.00 Page 10
LB-0265
4 System Specifications
C
EREBUS
S
TIM
S
WITCH
Stimulation Inputs 4 DB-37 front panel connectors
1.3 k impedance between input and output when stimulation
mode is enabled.
> 10 M input impedance when stimulation mode is disabled
Recording Outputs 36 Channel Samtec Connector with 32 output channels plus
reference and ground pins.
1.1 k impedance between input and output when read mode is
enabled.
> 10 M output impedance when read mode is disabled.
Minimum Switching Time < 1 µs when using the Enable In input to switch between modes
10ms when using StimComm to switch between modes
Maximum Time after switching
before action potentials are valid
< 5 ms using Fast Settle modified Cerebus Front End Amplifier
< 4 s using unmodified Cerebus Front End Amplifier
Maximum Stimulation Input
Current
30 mA on any one channel
Maximum Stimulation Input
Voltage Range
±15 V between any input and ground on any one channel
Enable In Input Voltage Range -0.1 V to +5.0 V
Bank to Bank Synchronization <1 µs skew.
PC Hardware Interface USB A-B cable
StimComm PC Software
Compatibility
Windows 2000 Pro, Windows XP Pro, Windows 7, or Windows
Vista
Internal Power Supply Standard 3-pin PC power connector accepting 110-240 VAC, 50-
60 Hz
Emergency Off Switch ¼” Mono Phone Plug
Normally Open Switch with < 1 k on resistance recommended.