Technical data

User’s and Service Guide 19
87050A Option K22 Commands
To connect all ports to their internal 50 Ω loads, send the following two commands:
OUTPUT 716;"PARAOUT52;"
OUTPUT 716;"PARAOUT105;" or
OUTPUT 712;"aterm"
OUTPUT 712;"bterm"
NOTE When a test set port is not in use, it is terminated in 50 Ω..
To read the Serial Number, send the following two commands:
OUTPUT 712;"sn?"
ENTER 712;S$
Reset Command:
When the *rst is set, the instrument is set to a known state where all ports are terminated.
B4 to PORT 10 b4p10
B4 to PORT 11 b4p11
B4 to PORT 12 b4p12
B4 to AUX2 b4aux
B1-4
TERMINATED
bterm 105 01101001
SERIAL
NUMBER
sn? Returns serial
number in a string
over the GP-IB
RESET
1
*rst
RESET2
2
*rst2
BOX IDENTITY idn? Displays on LCD
display only
1.
Previously, one reset command (*rst) resided in the firmware. See Figure 8 on
page 20
for switch and port configuration. This command resets all switches.
2. The new reset command (*rst2) resets only sw14, sw15, sw16, sw17 and sw50
through sw61. See
Figure 9 on page 20 for switch and port configuration.
Switches sw10 and sw13 do not change from their previous settings.
Table 2 Test Port Addresses
Connection Path GPIB Command Decimal Binary Equivalent