8. Click on Automation mode button and select ‘Sixpack use’ or ‘NMR case’
9. Select “1” for First sample
NOTE: ICON-NMR gives you the option to lock and shim the sample prior
to running the automation or, if you have stopped or halted the previous
automation run, to continue on the sample already in the magnet. For that
purpose click on the small button next to “First sample in the magnet (locked
and shimmed)?” and a check mark will appear. The system will skip the
inject/eject, locking and shimming for that sample. In addition the user can
choose to process all acquired data sets or do it later. This option can be set
in the ICON configuration set up.
10. Click on ‘Start’
NOTE: At this point the “ICON-NMR / Auto Online Controls” window and
the “acquisition status icons” appears.