Installation manual

86
have been completed or failed for experiment(s) that have been setup to run
on the sample.
5.2.2. Sixpack use and NMR case
NOTE: If the changer is not installed on top of magnet, remove any samples
from the magnet and then install the sample changer.
1. Turn the the switch on the sixpack/case controller box to OFF position
2. Turn on the sample lift (shim keypad) and remove the sample and leave
the lift air on
!!!WARNING!!!: Be absolutely sure that no sample is in the magnet and the
sample holder is empty before proceeding to the next step.
3. With the sample lift air still on, move the sample holder to the start
position (set the small hole over the shim stack)
4. Turn off the sample lift (from the keypad)
5. Load the samples into the sample holder
6. Turn the switch on the sixpack/case controller box to ON position
7. Click on ‘GO’
NOTE: The “Initialize Run” window appears.