needs two 1D-preparation experiments to optimize the 2D parameters in both
dimensions. In this case “PROTON” and “C13DEPT45” are used for the
optimization. Since the “PROTON” experiment already exists, only the
“C13DEPT45” is automatically entered into the setup on the present holder.
6. Click on the first line of Holder 1
7. Click on ‘Submit’
NOTE: The Status traffic lights of the four experiments change to yellow to
indicate the holder is ready to run and the experiment time of each
experiment is calculated. Proceed to “B. SAMPLE RUN” to start the
automation.
5.1.5. Set up of two or more holders with different solvents and
experiments
1. Follow steps 1 through 7 of “5.1.1. Setup of one single experiment”
2. Double click on Holder 2
3. Click inside the Name window and change the name to sample2
4. Click on the small button to the right of the solvent window and select
“DMSO” by clicking on it
5. Click on the small button to the right of the Experiment window and
select “C13CPD32” by clicking on it
6. Click on the small icon to the left of the Orig/Title window and type a
title