Installation manual

8
6. Click on ‘NEXT>’
7. Enter the RS232 ports for the external devices as shown in the picture
above.
NOTE: Use the default connection listed on the label on the inside of the
console. If a BACS 60 or 120 is used, select the proper RS232 port
(normally tty08) and be sure the power of the BACS is on. The sample
changer will configure the correct number of sample holders. If a SIXPACK,
CASE, MAS or HRMAS sample changer is used, set the BACS port to an
unused port number (for example tty20. After a few seconds a message will
appear that there is no communication to the sample changer and a default
of 60 sample holders is been used. Just click on the ‘OK’ button to continue
with cf. The number of sample holders for different sample changers can be
set in the ICONNMR configuration ( see 4.5.3 Default Number of Sample
Holders).