NOTE: “COSY45SW” is a composite experiment and is marked with the
letter C. A composite experiment contains two or more single experiments. In
this case, the “COSY45SW” is a 2D Homonuclear Correlation experiment
which needs a preparation experiment “PROTON” to prepare the “COSY
experiment to optimize the 2D parameters. The preparation experiment is
entered automatically into the setup on the present holder. The system will
check if a preparation experiment already exists on the present holder so that
it does not run the experiment twice.
3. Click on the small icon to the left of the Orig/Title window and type a
title
4. Click on the first line of Holder 1
5. Click on ‘Submit’
NOTE: The Status traffic lights of the two experiments change to yellow to
indicate the holder is ready to run and the experiment time of each
experiment is calculated. Proceed to “B. SAMPLE RUN” to start the
automation.