Installation manual
60
3.2.5.2. POST-INSERTION Set/Check
6. Click inside the square next to “Set & Check Temperature after Sample
Insertion”
7. Click on the arrow next to the “Temperature Setting after Sample
Insertion” window and select one of the following options:
• according to experiment’s TE Parameter: if selected, the
temperature value is been taken from the TE parameter
• ‘295K’: if selected, this value can be changed to the desired
temperature.
8. Click on the arrow next to the “Post-insertion Temperature Set/Check
Routine” window and select one of the following options:
• TESET; TEREADY 60 0.1: if selected, TESET sets the
temperature while TEREADY has 2 arguments which can be
changed. The first defines the maximum time in seconds to wait for
the temperature to be reached. The second parameter determine the