Installation manual
54
Loading”
7. Activate “Enable Shim file Loading” by clicking inside the small square
8. Select the probehead for which the shim file applies
NOTE: The current probehead has to be selected through edhead (Chapter
2.2). This probehead displayed as default in the “Probe Type to associate
shim files with” window.
9. Click on the folder icon next to the window of “Copy this shim file to all
entries for this Probe”
10. Select the shim file by clicking on it
11. Click on ‘Copy’
NOTE: The shim file has now been copied into all solvents. To load a shim
file only for a specific solvent, click on the folder icon next to the Associated
Shim File window for a specific solvent and select the shim file by clicking on
it.
Use standard shimfiles where available
If desired, solvent / probehead dependent shim files can be created. The shim
files have to be stored with the “wsh” command with the solvent name and the
probe ID as an extension (e.g. CDCl3.07). The files then have to be copied
from:
[TOPSPINHOME]/exp/stan/nmr/lists/bsms
into the corresponding prosol directories:
[TOPSPINHOME]/conf/instr/spect/prosol/[solvent.probe ID]
Once created, the option ‘Use Standard’ button can be used.
Lock -noauto
For some solvents or mixtures of solvents with several strong peaks, the lock