3.2.2.2. Automatic Shim File Loading
1. Click on “Automatic Shim File Loading
NOTE: When shimming on the sample is required, it is a good practice to
load a good shim file before locking. This applies for both the Automation
and Routine Spectroscopy. To activate this procedure, follow the steps below.
2. Go to TopSpin
3. Insert a sample into the magnet
NOTE: Use a clean sample of the same type which you do your research on.
Do not use any Standard Test Samples.
4. Lock and shim for best homogeneity
5. Type wsh and store the shim file. Use a short name (e.g. qnp5, bb5 or inv5
etc.)
6. Go back to ICON-NMR configuration and “Automatic Shim File