Installation manual

52
Shim after this number of samples
This option is activated when the setting in Shim the Sample option is set to
“after a solvent change or N samples” (see figure above) This is primarily used
for BEST NMR.
Always shim after a QNP change
If enabled, this option applies when the nucleus is changed on a QNP (quad
nucleus) probe.
Total time allocated for tune command (minutes)
This option applies for both the Automation and Routine Spectroscopy. The
setting is only active for the Automatic Shim Program “TUNESX, “TUNE
mytunefile” or “ROTATION-OFF;TUNESX;ROTATION-ON”.
Maximum Shim time (per iteration in minutes)
This option applies for both the Automation and Routine Spectroscopy. The
setting is only active with the Automatic Shim Program set to either
TUNESX”, “TUNE mytunefile” or “ROTATION-
OFF;TUNESX;ROTATION-ON”.