Installation manual
48
• Ask on Startup: in the Initialize Run Window a check box will
give the option to process the data after acquisition.
Automatic Lock Program
This option applies for both the Automation and Routine Spectroscopy. The
locking procedure in ICONNMR is controlled through the selections, LOCK
or SKIP LOKING. The options can be selected by clicking on the down arrow
button next to it.
Lock only after a solvent change
This option applies only in Automation and the use of a sample changer. It is
primarily for use with BEST NMR accessoriy.
Ignore the TOPSPIN Prosol Parameters
When enabled, ICON-NMR does not execute a pulsesort and therefore
ignores the prosol parameters. This option is primarily used for non standard
pulse programs and instrument configurations.
Never Rotate the Sample
When enabled, all sample are run non spinning. This option applies for both