Mode
By clicking on the icon, the mode can be changed. The sun icon stands
for “anytime” and the moon fornight time” operation. To make use of
this feature, the Priority and Night/Day switch over has to be enabled
and the switch over time has to be set (see B.4. Priority in this Chapter).
Each user can be setup to have their own customized day and night
time experiments. This way the users may run experiments only at a
specific time
Details
The letter N in front the experiment name stands for Normal which
denotes a single or stand-alone experiment. The parameters for such an
experiment are stored in the “TOPSPINHOME/exp/stan/nmr/par/”
directory under the same name. To add a new user defined N-
experiment and parameter set, see Chapter 7, section 7.2 Adding the
new experiment to ICON-NMR.
The letter C in front of the experiment name stands for Composite
experiment and represents a series of two or more experiments
combined under one name. To add a new C-experiment, see Chapter 7,
section 7.3 adding a new composite experiment.
3.1.1.3. Data Set Names box
Predefined data set names can be assigned to the different users. The name(s)
can then be chosen from a list during the setup of the experiments in Routine
Spectroscopy and Automation.
1. Click inside the Data Set Name box and an Entry Box window appears