Installation manual
21
14. Select nucleus: ‘29Si’, change and Calculate the following parameters:
Standard parameters for 29Si on channel F2 routed to amplifierA2
90 deg. decoupler pulse length same as 90 deg.
transmitter
power level same as transmitter
cpd pulse length 100
power level click on ‘calc’
15. Click on ‘Save’
16. Select nucleus: ‘31P’, change and Calculate the following parameters:
Standard parameters for 31P on channel F2 routed to amplifierA2
90 deg. decoupler pulse length same as 90 deg.
transmitter
power level same as transmitter
cpd pulse length 63
power level click on ‘calc’
17. Select logical channel ‘Global’ by clicking on it
18. Change the following special parameters:
grad. recovery delay 0.0005 (only if equipped with
gradients)
grad. pulse 1 1000 (only if equipped with
gradients)
grad. pulse 2 1000 (only if equipped with
gradients)
trim pulse mlev 2500
trim pulse hsqc 2000
19. Select ‘Standard soft pulses by clicking on it