2.3. Prosol and TUNE file set up
1. Type edprosol
NOTE: To follow the instruction below, the transmitter 90 deg. Pulse length
and power levels of the observe nuclei have to be known and entered in the
appropriate fields. If ATP was used to calibrate the probe, those parameters
were automatically entered. If the transmitter 90 deg. Pulse length and
power levels are not known, then use the instructions in Chapter 3,
Optimizing Parameters, to measure them. It is a good practice to use this
chapter to measure the exact pulse length and power levels. Choose only
those nuclei which can be observed on the current probe.
2. Select logical channel ‘F1 + F2’ by clicking on it
3. Select ‘Standard hard pulses’ by clicking on it
4. Select Probe name: ‘current’
5. Select solvents: ‘all’