Specifications

PRODUCT SHEET
info@biopac.com
support@biopac.com
www.biopac.com
BIOPAC Hardware | MP150 Systems | Page 2 - 9 Updated: 2.25.2015
MP150 SYSTEM SPECIFICATIONS
Analog Inputs
Number of Channels: 16
Absolute Maximum Input: ±15 V
Application Programming Interfaces options:
Hardware Interface BHAPI
Software Interface ACKAPI
Operational Input Voltage:
±10 V
A/D Resolution: 16 Bits
Accuracy (% of FSR): ±0.003
Input impedance: 1.0 M
Analog Outputs
Number of Channels: 2
Max output with acquisition: 2 channels
Output Voltage Range: ±10 V
D/A Resolution: 16 bits
Accuracy (% of FSR): ±0.003
Output Drive Current: ±5 mA (max)
Output Impedance: 100
Digital I/O
Number of Channels: 16
Voltage Levels: TTL, CMOS
Digital I/O Logic Type: CMOS
Input Voltage Range: -0.5 V to 5.5 V (max)
Input Clamp Current: ±20 mA (max)
Output Drive Current: ±20 mA (max)
External Trigger Input: TTL, CMOS compatible - See also: External Trigger Inputs
Logic Level Thresholds:
Input Low Voltage: 1.50 V (max)
Input High Voltage: 3.45 V (min)
Time Base
Min Sample Rate: 2 samples/hour
Trigger Options: Internal, External or Signal Level
Power
Amplifier Module Isolation: Provided by the MP unit, isolated clean power
CE Marking: EC Low Voltage and EMC Directives
Leakage current: <8 µA (Normal), <400 µA (Single Fault)
Fuse: 2 A (fast blow)
Device specs
MP150
Max Sample Rate
MP Internal Memory:
200 K samples/sec (400 K aggregate)
PC Memory/Disk: 200 K samples/sec (400 K aggregate)
Internal Buffer: 6 M samples