Instruction manual

Volts. The input VOLTAGE data is in Numeric Data for-
mat. The <Channel_list> above, indicates for which all
CHANNELS the DATA has to be SET. The same data will
be set on all the specified CHANNEL(S). If no such
<Channel_list> is provided the SOURce command will
always SET to channel 1.
For example to set 10.5 Volts on Channel 2, send the
command as:
SOURce:VOLTage 10.50(@2)
For example to set 30 Volts on Channels 1&2,send the
command as: VOLTage 3.00 E+1(@1:2)
3. Command : [SOURce:]CURRent<SP><NR.DATA>
<Channel_list>
Description : Set the Output Current Limit for that Channel(s).
Function : This command sets the Output Current of the Power
Supply output with the <NR.DATA> given in the command
in AMPS. The input CURRENT data is in Numeric Data
format. The <Channel_list> above, indicates for which
all CHANNELS the DATA has to be SET. The same data
will be set on all the specified CHANNEL(S). If no such
<Channel_list> is provided the SOURce command will
always SET to channel 1.
For example to set 2.3 Amps on Channels 1&2 send the
command as:
SOURce: CURRent 2.300 E+0(@1,2)
For example to set 1 Amp on Channel 1 only, send the
command as:
CURR 1
For example to set 1.99 Amps on Channel 2, send the
command as:
SOUR:CURR 199 E-2(@2)
4. Command : [SOURce:]VOLTage:LIMit<SP>
<NR.DATA><Channel_list>
Description : Set the Upper Limit of O/P Voltage for that Channel(s).
Function : This command sets the Soft Upper Limit for Output
Voltage of the specified CHANNEL(S) of the Power
Supply output with the <NR.DATA> given in Volts. The
input VOLTAGE data is in Numeric Data format. The
<Channel_list> above,
53