Figure 8, which prompts for a “short”. Also, “REL” 
indicator will be blinking on the display. This is to 
prompt the user that the meter is waiting for an action. 
At this point, short calibration will be done when the 
 button is pressed once. Within a few seconds, 
the meter will return to normal display and meter will 
be calibrated based on the test frequency and settings 
selected prior to entering the calibration mode. 
Figure 8 - Short Calibration 
Quick Procedure 
Below is an example of steps to follow to do both 
open and short calibration. 
1.  Select the primary function mode for 
measurement (i.e. L/C/R/Z). 
2.  Select the test frequency for measurement. 