User manual
Table Of Contents
- 1.1 Feature Overview
- 1.2 Input Power and Fuse Requirements
- 1.3 Package Contents
- 2.1 Front Panel Overview
- 2.2 Screen Display
- 2.3 Front Panel Menu Options
- 2.4 Front Panel Menu Overview
- Rear Panel Summary
- 2.6 Power up
- 3.1 Overview
- 3.2 Measuring Voltage
- 3.3 Measuring Current
- 3.4 Measuring Resistance
- 3.5 Measuring Frequency and Period
- 3.6 Measuring Continuity
- 3.7 Testing Diode
- 3.8 Math Functions
- 4.1 Measurement configuration
- 4.2 Trigger Operations
- 4.3 Buffer Operations
- 4.4 Limit Operations
- 4.5 System Operations
- 5.1 Selecting an Interface
- 5.2 USB & RS-232 Interface Operation
- 5.3 GPIB Interface operation (model 5492BGPIB only)
- 5.4 Data Format
- 6.1 Command Structure
- 6.2 Command Syntax
- 6.3 Command Reference
- 7.1 Frequently Asked Questions
- 7.2 Error Messages
- 8.1 Technical Specifications

SCPI Command Reference
85
Frequency and Period) is set using this :NPLCycles command. NPLC (Number of Power Line
Cycles) expresses the integration period by basing it on the power line frequency. For example, for
a PLC of 1, the integration period in seconds would be1/60 (for 60Hz line power) or 16.67 ms.
Note: Set the measurement function first before setting or querying the NPLC. All the
command syntax listed will set or query the same NPLC value.
RANGe Command
The RANGe command is a subset to current, voltage, and resistance measurement commands, of
which are part of the SENSe subsystem. The below lists out all the command syntax for setting RANGe
for different measurement functions.
:RANGe:[UPPer] <n>
Command syntax:
[:SENSe[1]]:CURRent:AC:RANGe[:UPPer] <n>
[:SENSe[1]]:CURRent:DC:RANGe[:UPPer] <n>
[:SENSe[1]]:VOLTage:AC:RANGe[:UPPer] <n>
[:SENSe[1]]:VOLTage:DC:RANGe[:UPPer] <n>
[:SENSe[1]]:RESistance:RANGe[:UPPer] <n>
[:SENSe[1]]:FRESistance:RANGe[:UPPer] <n>
Command Parameter:
<n> = 0.01 (12 mA), 0.1 (120 mA), 1 (1.2 mA), 10 (12 A) For DCI
0.01 (12 mA), 1 (1.2 mA), 10 (12 A) For ACI
0.1 (120 mV), 1 (1.2 V), 10 (12 V), 100 (120 V), 750 (750 V) For ACV
0.1 (120 mV), 1 (1.2 V), 10 (12 V), 100 (120 V), 1000 (1000 V) For DCV
100 (120 Ω), 1000 (1.2 KΩ), 10000 (12 KΩ), 100000 (120 KΩ),
1000000 (1.2 MΩ), 10000000 (12 MΩ), 100000000 (120 MΩ) For Ω2/Ω4 Wire
DEFault 10 (12 A range: ACI and DCI)
750 (750 V range: ACV)
1000 (1000 V range: DCV)
100e6 (120 MΩ range: Ω2/Ω4 Wire)
MINimum 0 (All functions)
MAXimum Same as DEFault
Query:
:RANGe[:UPPer]? Query measurement range of the currently selected function.
Description:
This command is used to manually select the measurement range for the specified measurement
function.