User manual

15
Sine Menu
F1: FrequencySelects and displays the frequency. Change the frequency setting using the
cursor keys, rotary knob, or numerical keys.
F2: AmplitudeSelects and displays the amplitude. Change the amplitude setting using the
cursor keys, rotary knob, or numerical keys.
F3: Offset Selects and displays the offset parameter. Change the offset by using the cursor
keys, rotary knob, or numerical keys. If a certain setting cannot be produced, the waveform
generator will display a Setting Conflict” message.
Amplitude and offset settings interact and are bound by hardware restrictions. To obtain the
desired waveform, the amplitude and offset must satisfy the following formula:
(Vp-p)/2 + |offset| <= 5 volts
F4: SymmetryWhen the Square or Triangle waveforms are selected, the SYMMETRY
(duty cycle) parameter is available. Change the symmetry (Triangle) or duty cycle (Square) by
using the cursor keys, rotary knob, or numerical keys. If a certain setting cannot be produced,
the waveform generator will display a warning message.