User manual

46
SWEep
This command activates or deactivates sweep:
Arguments
Type: Boolean
Command Type: Setting or Query
Setting
Syntax: SWE<ws>ON|OFF|LIN|LOG
Examples: SWE ON
SWE LIN
Query
Syntax: SWE?
Response: OFF|LIN ON|LIN OFF|LOG ON|LOG OFF
Note: Sweep will automatically be active if set to Linear or Logarithmic.
SWRAte
This command sets the time for one complete sweep:
Arguments
Type: Numeric
Units: S, mS, uS, nS
Range: 10mS to 100S
Command Type: Setting or Query
Setting
Syntax: SWRAte<ws><time>[units]
SWRAte<ws>MINimum|MAXimum
Examples: SWRAte 50MS
Query
Syntax: SWRAte?[<ws>MINimum|MAXimum]
Response: NR3
SWSTArt
This command sets the start frequency of the sweep:
Arguments
Type: Numeric.
Units: MHz, KHz, Hz (default)
Range: Dependent on the frequency range of the current function.
Command Type: Setting or Query
Setting
Syntax: SWSTArt<ws><frequency>[units]