User manual
46 
SWEep 
This command activates or deactivates sweep: 
Arguments     
Type:      Boolean     
Command Type: Setting or Query   
Setting      
Syntax:     SWE<ws>ON|OFF|LIN|LOG  
Examples:  SWE ON   
  SWE LIN   
Query     
Syntax:     SWE?   
Response:   OFF|LIN ON|LIN OFF|LOG ON|LOG OFF    
Note: Sweep will automatically be active if set to Linear or Logarithmic. 
SWRAte 
This command sets the time for one complete sweep: 
Arguments     
Type:      Numeric   
Units:     S, mS, uS, nS    
Range:     10mS to 100S       
Command Type: Setting or Query   
Setting      
Syntax:      SWRAte<ws><time>[units]  
     SWRAte<ws>MINimum|MAXimum    
Examples:     SWRAte 50MS   
Query     
Syntax:     SWRAte?[<ws>MINimum|MAXimum]   
Response:    NR3   
SWSTArt 
This command sets the start frequency of the sweep: 
Arguments     
Type:      Numeric.    
Units:     MHz, KHz, Hz (default)  
Range:      Dependent on the frequency range of the current function.      
Command Type: Setting or Query   
Setting      
Syntax:     SWSTArt<ws><frequency>[units]  










