Instruction manual

69
SCPI commands overview
DISPlay Subsystem
This subsystem controls the presentation of textual information and
measurement data.
:DISPlay[:WINDow][:STATe] {0|1|OFF|ON}
This command turns power supply’s display off and on.
When the display is off, only indicators are displayed.
When the display is on, power supply is in nomal mode (measured or
programmed values of voltage and current are displayed.).
You can replace off|on parameters with 0|1 numeric values.
:DISPlay[:WINDow][:STATe]?
This command queries front panel display status. It returns only numeric
values: 0 (off) or 1 (on).
:DISPlay[:WINDow]:TEXT[:Data] <quoted string>
This command displays a message on the front panel. The power supply will
display up to 16 characters and all the rest will be truncated. The string can be
sent between simple or double quotes.
:DISPlay[:WINDow]:TEXT[:DATA]?
This command queries the displayed message. It returns a quoted string of
maximum 16 characters.
:DISPlay[:WINDow]:TEXT:CLEar
This command clears the message displayed on the front pannel. Only
indicators will be displayed