Instruction manual

65
SCPI Command Terminators
A command string sent to the power supply must terminate with a new line
character (ASCII decimal code of 10) or a a carriage return character (ASCII
decimal code of 13).
For multiple commands sent in a single message, command separator is
semicolon character (03b h).
Important note!
The power supply will go to remote mode of operation (it can accept
commands over RS-232 interface) if SYSTem:REMote command is sent.
If other remote interface commands are sent before sending SYSTem:REMote
command, the power supply will respond with Power supply in local mode
message.