Instruction manual
27
RS-232 Configuration
Available settings for RS-232 interface:
♦ Baud rate: 1200, 2400, 4800, 9600 (factory setting: 9600)
♦ Parity and data bits: None – 8 data bits (factory setting)
Odd – 7 data bits
Even – 7 data bits
♦ Number of start bits: 1 bit (cannot be changed)
♦ Number of stop bits: 1 bit (cannot be changed)
Set RS – 232 remote interface parameters
1. Select RS – 232 interface
Press Remote / Local key. If “GPIB / 488” is displayed, turn the knob until
“RS-232” will be displayed.
Press Remote / Local key again. Baud rate settings will be displayed.
2. Select baud rate
By turning the knob, you will view available baud rates: 1200, 2400, 4800,
9600 bits per second.
When desired baud rate is displayed, press Remote / Local key. Parity settings
will be displayed.
3. Select parity
By turning the knob, available parities will be displayed:. Even, odd or no
parity. Select the desired parity.
By pressing Remote / Local key, the settings will be saved in the non-volatile
memory and Saved message will be displayed.
If you didn’t change any of the previous set parameters, the power supply will
display No Change message.
If you did change a single parameter from the previous set parameters, Saved
message will be displayed.
After one of these messages is displayed for several seconds, the power supply
returns to the previous state.
RS - 232