Error 611: EEPROM absent
Error 612: ADC offset register V checksum failed
Error 613: ADC offset register I checksum failed
Error 614: DAC offset register V checksum failed
Error 615: DAC offset register I checksum failed
Error 616: ADC full gain register V checksum failed
Error 617: ADC full gain register I checksum failed
Error 618: DAC offset register OVP checksum failed
Error 619: DAC V step register checksum failed
Error 620: DAC I step register checksum failed
Error 621: DAC OVP step register checksum failed
If one of the errors above is generated, the calibration procedure must be done.
If one of these errors is generated again, the power supply must be delivered to
B&K Precision for service
.
Error 630: Data in location 1 checksum failed
If this error is generated, a new saving for the power_up state must be done. If
the error persists, the power supply must be delivered to B&K Precision for
service.
Error 641: Filter register error
Error 642: ADC not locked
Error 643: ADC not ready
Error 644: ADC out of range
Error 651: DAC out of range
Error 661: I/O processor does not respond
Error 662: I/O processor failed self test
Error 663: I/O processor communication error