Return Meter 
This option allows users to enable an internal fixed timer delay (5 seconds) for the power supply 
to automatically switch from setting display to measured display. When enabled, if the power 
supply output state is ON (enabled) and if the display shows setting voltage and current, it will 
automatically switch to measured voltage and current display after 5 seconds. The backlight of 
the   button will also be lit. In this state, pushing the   again to toggle back to 
setting display will reset the 5-second timer again before the supply changes back to measured 
display. 
To configure this option, follow the steps below: 
1.  From the SYSTEM menu, browse and select ReturnMeter and press  . 
2.  Select between the two options: 
Off(Def) – Return Meter function disabled. 
Delay – Enables return meter function. 
3.  Select the desired option and press   to save the change. 
4.  To exit the menu at any time, press   twice. 
3.4  CONFIG Menu 
All setup procedures and settings explained in this section can be accessed from the CONFIG 
menu. To access this menu, press   and press   . Select CONFIG, then press 
. 
Load Setup Option 
The power supply has an internal dummy load that can be enabled to increase the speed of the 
voltage fall time for high speed test applications. The effectiveness of this function is 
dependent on the DUT (device under test) and may or may not be useful for some applications.  
This feature should only be used with caution, as it is not designed for all applications. 
DO NOT enable this function for applications such as connecting devices for battery charging 
or powering electric motors. All other applications that may behave similarly as a load should 
NOT use this function. This function is disabled by default. 