Restore Factory Default Settings 
All instrument settings can be reset back to their factory default values by doing the following: 
WARNING: 
Restoring the instrument to factory default will change all current instrument 
settings and parameters back to their default values. 
1.  From the SYSTEM menu, select Initialize and press  . 
2.  The instrument will return to the normal display and all settings are now restored back 
to factory default. The below table lists some of the factory default settings. 
Table 2 - Factory Default Settings 
Communication Address 
0 
Key Sound 
ON 
Trigger Mode 
Manual 
Communication Interface 
RS232 
RS232 Settings 
4800, 8, N, 1 
Return Meter 
Off 
Memory 
Group 0 
Power-On 
Rst 
Load Setup 
Off 
External Control 
10v-M 
Limit 
Vmin = 0.000V, Vmax = 81.000V 
Online Setup 
Off 
Output 
Disabled 
Configure Power-On State 
The initial power-On state of the power supply can be configured by following the steps below: 
1.  From the SYSTEM menu, select Power-On and press  . 
2.  There are two options: 
Rst(Def) – Factory Default. 
Sav0 – Settings before last power up. 
3.  Select the settings you want during power up, and press   to save changes. 