User manual
17 
WARNING:
Before connecting wires to the output terminals, turn 
OFF the power supply to avoid damage to the 
instrument and the device under test (DUT). For 
safety, load wires must have a wire gauge size large 
enough to prevent overheating when the power 
supply operates at maximum short circuit output 
current. It will also prevent large voltage drops from 
resistances in the wires. 
2.3  Preliminary Check 
Complete the following steps to verify that the power supply is ready for use. 
1.  Verify AC Input Voltage 
Verify and check to make sure proper AC voltages are available to power the instrument. 
The AC voltage range must meet the acceptable specification as explained in “2.1 Input 
Power and Fuse Requirements”. 
2.  Connect Power and Self Test 
Connect AC power cord to the AC receptacle in the rear panel and press the power 
switch to the |(ON) position to turn ON the instrument.  It will run through a self test 
procedure with the screen shown in below: 
Self-test Errors 
The following errors will be displayed if self-test did not complete successfully: 
Error Message on Display 
Description 
EEPROM FAILURE 
The internal EEPROM is corrupted or damaged. 
Config Data Lost 
The last operation data within the EEPROM is lost. 
Calibration Data Lost 
Calibration data within the EEPROM is lost. 
FactoryCal.Data Lost 
Factory calibration data is lost. 
MainframeInitialize Lost 
The system settings within the EEPROM is lost. 
System Selftest . . . . 
. . 










