SWEep:TIME <time> 
Set sweep time. 
SWEep:TIME? 
Check the sweep time. 
SWEep:SOURce {INTernal|EXTernal} 
Set trigger source of sweep as internal (INTernal) or external (EXTernal).  
SWEep:SOURce? 
Check current trigger source. 
SWEep:STATe {ON|OFF} 
Turn on sweep mode (ON) or turn off sweep mode (OFF). 
SWEep:STATe? 
Check the “on/off” state of sweep mode. The returned data are “0” (OFF) or “1” (ON). It returns to 
fixed frequency mode in “off” state. 
Counting instructions 
FUNCtion:TOTal INITial 
Initialize the counter. 
FUNCtion:TOTal STARt 
Set the counter to start counting. 
FUNCtion:TOTal STOP 
Set the counter to stop counting. 
FUNCtion:TOTal CLEAr 
Clear the counted value of the counter. 
FUNCtion:TOTal? 
Record the current value of the counter. 
Frequency Measurement instructions 
FUNCtion:FREQuency MEASure 
Enable frequency measurement function for the instrument to start a new frequency measurement. 
FUNCtion:FREQuency? 
Record last frequency value after frequency measurement is finished. 
FUNCtion:FREQuency GATE <time> 
Set gate time in the range of 10ms to 10s for frequency measurement. 
FUNCtion:FREQuency GATE?  
Check current gate time for frequency meadurement. 
Trigger instructions 
TRIGger:SOURce {IMMediate|EXTernal|BUS} 
Set trigger source or modulating source for available functions as internal (IMMediate), external 
(EXTernal) or single (BUS). 
TRIGger:SOURce? 
 Check trigger source or modulating source of current function as internal (IMM), external, (EXT) 
or single (BUS). 