Instruction manual

6
Section 3
Technical Specifications
3.1 FREQUENCY CHARACTERISTICS
Waveforms: Sine, Square, Triangle, Ramp, ± Pulse
Range: 0.5 Hz to 4 MHz in 6 ranges
Tuning Range: Variable multiplication factor of 0.04 to 4.0 of selected range
Resolution: 0.001 Hz (model 4003A only)
Operating Modes: Normal, Sweep, VCF
Frequency Stability: Output will change less than 0.1 % for 15 minutes after switching ON
and it will change less than 0.2 % for 24 Hours after switching ON
3.2 OUTPUT CHARACTERISTICS
Impedance: 50 ± 2 %.
Level: Variable control
From
0.2 Vpp (no load, Amplitude knob in pull position)
0.1 Vpp (50 load, Amplitude knob in pull position)
To
20 Vpp (no load, Amplitude knob in push position)
10 Vpp (50 load, Amplitude knob in push position)
Attenuation: -20 dB ±2%
DC Offset: Variable: ±10 V open circuit,
± 5 V into 50
3.3 SINE WAVE
Distortion: < 2% (1 Hz 100 kHz)
Harmonic Ratio: < 30 dB, 100 kHz to 4 MHz
Frequency Response: < 0.5 dB to 100 kHz
< 1.5 dB, 100 kHz to 4 MHz
3.4 TRIANGULAR WAVE
Symmetry: 50 % (positive half) to 50 % (negative half)
< 2%, 1 Hz to 100 kHz
3.5 RAMP WAVE
Frequency Range: 0.5 Hz to 3.5 MHz
Symmetry: 80% (rise wave) to 20% (fall wave)
< 5%, 1 Hz to 100 kHz
Rising Wave Linearity: < 2%, 1 Hz to 100 kHz