Instruction manual
3
1. Introduction 
1.1 General 
The B&K Precision Corp. 889B Synthesized In-Circuit LCR/ESR Meter is a high accuracy test instrument used 
for measuring inductors, capacitors and resistors with a basic accuracy of 0.1%. Also, with the built-in functions of 
DC/AC Voltage/Current measurements and Diode/Audible Continuity checks, the 889B can not only help 
engineers and students to understand the characteristics of electronics components but also being an essential 
tool on any service bench. 
The 889B is defaulted to auto ranging. However, it can be set to auto or manual ranging by pressing the Range 
Hold key. When LCR measurement mode is selected, one of the test frequencies, 100 Hz, 120 Hz, 1 KHz, 10 
KHz, 100 KHz or 200 KHz, may be selected on all applicable ranges. One of the test voltages, 50mVrms, 0.25 
Vrms, 1 Vrms or 1 VDC (DCR only), may also be selected on all applicable ranges. The dual display feature 
permits simultaneous measurements. When DC/AC voltage/current measurement mode or the Diode/Audible 
Continuity Check mode is selected, only the secondary display will be used to show the result of the 
measurement. 
The highly versatile 889B can perform virtually all the functions of most bench type LCR bridges. With a basic 
accuracy of 0.1%, this economical LCR meter may be adequately substituted for a more expensive LCR bridge in 
many situations. Also, with the basic accuracy of 0.4% in voltage and current measurements, the 889B performs 
the functions of a general purpose Digital Multi-Meter and can be used to replace the DMM on a service bench. 
The 889B has applications in electronic engineering labs, production facilities, service shops, and schools. It can 
be used to check ESR values of capacitors, sort and/or select components, measure unmarked and unknown 
components, and measure capacitance, inductance, or resistance of cables, switches, circuit board foils, etc. 
The key features are as following: 
1. Voltage Measurements: 
  AC  : True RMS, up to 600Vrms @ 40 ~ 1K Hz 
  DC  : up to 600V 
  Input Impedance  : 1M-Ohm 
2. Current Measurements: 
  AC  : True RMS, up to 2Arms @ 40 ~ 1K Hz 
  DC  : up to 2A 
  Current Shunt  : 0.1 Ohm @  20mA; 10 Ohm @  20mA 
3. Diode/Audible Continuity Checks: 
  Open Circuit Voltage: 5Vdc 
  Short Circuit Current: 2.5mA 
 Beep On:  25  
 Beep Off:  50  
4. LCR Measurements: 
  Test conditions 
  Frequency  :  100Hz / 120Hz / 1KHz / 10KHz / 100KHz / 200KHz 
  Level  :  1Vrms / 0.25Vrms / 50mVrms / 1VDC (DCR only) 
  Measurement Parameters  : Z, Ls, Lp, Cs, Cp, DCR, ESR, D, Q and θ 
 Basic Accuracy  :  0.1% 
  Dual Liquid Crystal Display 
  Auto Range or Range Hold 
 USB Interface Communication 
 Open/Short Calibration 
 Primary Parameters Display: 
Z : AC Impedance 
DCR : DC Resistance 
Ls : Serial Inductance 
Lp : Parallel Inductance 