Instruction manual
9
Z Accuracy: 
  As shown in table 1. 
C Accuracy: 
Cxf
Zx
2
1
C
Ae
 = Ae of C 
f  :  Test Frequency (Hz) 
Cx  :  Measured Capacitance Value (F) 
|Zx|  :  Measured Impedance Value () 
Accuracy applies when Dx (measured D value)  0.1 
When Dx > 0.1, multiply C
Ae
 by 
2
1 Dx  
Example: 
Test Condition: 
 Frequency  : 1KHz 
 Level  : 1Vrms 
 DUT  : 100nF 
Then 
1590
9
10100
3
102
1
2
1
Cxf
Zx
Refer to the accuracy table, get C
Ae
=±0.1% 
L Accuracy: 
LxfZx 
2
L
Ae
 = Ae of L 
f  :  Test Frequency (Hz) 
Lx  :  Measured Inductance Value (H) 
|Zx|  :  Measured Impedance Value () 
Accuracy applies when Dx (measured D value)  0.1 
When Dx > 0.1, multiply L
Ae
 by 
2
1 Dx
Example: 
Test Condition: 
 Frequency  : 1KHz 
 Level  : 1Vrms 
 DUT  : 1mH 
Then 
283.6
3
10
3
102
2
LxfZx
Refer to the accuracy table, get L
Ae 
= ±0.5% 