Figure 2 Calibrating the Site Analyzer
Step Action
1 Press the Calibrate key
2 Connect the Cal Combo “Open” connector to
the unit antenna port
3 Press the Open softkey then wait for the
unit to beep and display Done
4 Connect the Cal Combo “Short” connector to
the unit antenna port
5 Press the Short softkey then wait for the
unit to beep and display Done
6 Connect the Cal Combo “Load” connector to
the unit antenna port
7 Press the Load softkey then wait for the
unit to beep and display Done
8 Remove the Cal Combo from the unit