User Manual

Table Of Contents
18
The device can monitor the heart rhythm for at least 3 more months.
At least 6 maximum energy shocks can be delivered until EOS occurs.
The selected parameters in the device program do not change.
EOS replacement indication
End of Service can be detected by Home Monitoring.
VT and VF detection and all therapies are deactivated!
The antibradycardia function remains active in the VVI mode:
Ventricular pacing: RV; basic rate 50 bpm; without special pacemaker functions
such as hysteresis, etc.
Pulse amplitude of 6 V; pulse width of 1.5 ms
Time of transmission for Home Monitoring: 90 days
Explantation and Device Replacement
Explantation
Interrogate the device status.
Deactivate VT and VF therapies prior to explantation.
Remove the leads from the header. Do not simply cut them loose.
Use state-of-the-art techniques to remove the device and, if necessary, the leads.
Explants are biologically contaminated and must be disposed of safely due to risk of
infection.
Device replacement
If, upon replacing the device, already implanted leads are no longer used but left in the
patient, then an additional uncontrolled current path to the heart can result.
Deactivate VT and VF therapies prior to device replacement.
Insulate connections that are not used.
Basic principles:
The device must not be resterilized and reused.
Cremation
Devices must not be cremated.
Explant the device before the cremation of a deceased patient.
Disposal
BIOTRONIK takes back used products for the purpose of environmentally safe disposal.
Clean the explant with an at least 1% sodium hypochlorite solution.
Rinse off with water.
W
CAUTION
Temporally limited therapy
If ERI occurs shortly after follow-up and is only detected during the subsequent
follow-up, then the remaining service time can be much less than 3 months.
Replace device soon.
W
WARNING
Patient at risk of death
If EOS replacement indication occurs before replacement of the device, then the
patient is without therapy.
Replace device immediately.
Monitor patient constantly until immediate replacement of the device!
Note:
Normal oxidation processes may cause ICD housing discolorations. This is
neither a device defect nor does it influence device functionality.