User Manual

43 Using Renamic
ECG and IEGM Functions
Connecting the ECG cable See section: Connect ECG cable, p. 23
ECG recorder and ECG
monitor
All ECGs can be displayed in real time in the recorder or trigger mode and printed
on the internal printer.
ECG leads
Tabelle 4: Accessories for ECG leads
See also: Accessories, p. 49
Recording ECGs and IEGMs The intracardiac electrograms received from the implanted device and the surface
ECG can be displayed and printed simultaneously. The recording of the surface ECG
is independent of other functions, so that the implanted device can be interrogated
and programmed during the ongoing ECG display. The recorded electrograms can
be saved and measured using electronic calipers.
Overmodulated ECG input When the ECG input is overmodulated, the signal is displayed only as a solid line on
the upper frame of the ECG window. Proceed as follows to fix overmodulation:
Lead Patient Device
Up to 3 leads (Einthoven) Permitted adhesive and clamp
leads
PK-222
Note: Follow the instructions for connecting the ECG cable PK-222 in section
Connect ECG cable, p. 23.
Note: It is not permitted to use the device's ECG display for diagnostic purposes
because it does not meet all requirements for diagnostic ECG devices (IEC 60601-
2-25).
Step Description
1 Test the lead contacts.
2 Remove other devices from the patient.
3 Turn off the sources of interference.