34 Using Renamic
Location of the emergency
buttons
Bild 3: Location of the emergency buttons on the device
Start emergency pacing Proceed as follows:
Stop emergency pacing Proceed as follows:
Trigger emergency shock Proceed as follows:
Step Action
1 Position the PGH above the implanted device so that telemetry contact
is created.
2 Press the safe program button:
Step Action
1 Activate the desired permanent parameters in the program of the
implanted device and transfer them to the implanted device.
Step Action Result
1 Position the PGH above
the ICD so that telemetry
contact is created.
Telemetry contact is created between the
ICD and the Renamic programmer.
2 Press the emergency
shock button:
The emergency shock parameters are
activated.
As a safety precaution, a dialog
window will give you the option to
cancel this action.
3 In the dialog window,
select [EMERGENCY
SHOCK].
The ICD shock capacitors are charged.
Renamic triggers a 30 or 36 J emer-
gency shock via the ICD.