User Manual
21 Startup
Transportation and Setup
Transporting the device • Renamic has an integrated ergonomic handle in the front and a gripping tab in
the back, which can be used to safely transport the device in any position.
• A carrying strap can also be attached to the device.
• The specially designed anti-slip pads allow for horizontal or vertical positioning
of the device.
• When the device is slightly lifted (using the handle), the slick corners of the base
allow for easy positioning on smooth surfaces (tables, shelves).
• After setting the device down, the anti-slip pads keep the device securely in
place.
Setting up the device
• Place the device on a flat dry surface. Make sure that it cannot shift even with
the cable connected and that the patient can only come into contact with the
applied parts, namely the programming head and ECG cable. The physician
must not touch any connections such as USB ports or interfaces for modules or
the programming head and the patient at the same time.
Tilting the screen up • In transport position (screen closed), unlock the screen by pressing both
release buttons at the same time. You can hear and feel the device unlock.
• Hold the sides of the screen with both hands and tilt it up to the position you
would like to use it in (1).
• Pivot the screen around the upper end of the screen arm (2.). The operating
position can be smoothly adjusted as needed.
The screen will remain in any position due to its self-retaining bearings.
WARNING
Danger to the user
Danger of tripping over connected cables during device transport.
• Prior to transporting the device, remove the attached cables and store them
in the compartments intended for this purpose.
WARNING
Danger to the patient
The device is not sterile and cannot be sterilized.
• Do not set up the device in a sterile area and do not position the device so that
the fan blows air into a sterile area.
Note: The device can be operated in the patient's environment.