User Manual

Table Of Contents
30
Parameters
Pacing and Sensing
Pacing and Sensing
Pulse amplitude and
pulse width
Sensitivity
Atrial capture control
Ventricular capture control
Atrial overdrive pacing
Parameter Range of values Standard SR DR HF
Pulse
amplitude A/RV/LV
0.2 ... (0.2) ... 6.0 ... (0.5)
... 7.5 V
3.0 V x x x
Pulse width A/RV/LV 0.1 ...(0.1) ... 0.5 ... (0.25)
... 1.5 ms
0.4 ms x x x
Parameter Range of values Standard SR DR HF
Sensitivity AUTO; 0.5 ... (0.5) ... 7.5 mV AUTO x
Sensitivity A AUTO; 0.1 ... (0.1) ... 1.5 ...
(0.5) ... 7.5 mV
AUTO x x
RV sensitivity AUTO; 0.5 ... (0.5) ... 7.5 mV AUTO x x x
LV sensitivity OFF; AUTO; 0.5 ... (0.5) ...
7.5 mV
AUTO x
Parameter Range of values Standard SR DR HF
Atrial capture control ATM (monitoring only); ON;
OFF
ON x x
Minimum amplitude 0.5 ... (0.1) ... 4.8 V 1.0 V x x
Threshold test start 2.4 ... (0.6) ... 4.8 V 3.0 V x x
Safety margin 0.5 ... (0.1) ... 1.2 V 1.0 V x x
Search type Interval; time of day Time of day x x
Interval 0.1; 0.3; 1; 3; 6; 12; 24 h 24 h x x
Time of day 00:00 ... (00:10) ...
23:50 hh:mm
00:30 hh:mm x x
Parameter Range of values Standard SR DR HF
Capture control RV ATM (monitoring only); ON;
OFF
ON x x x
Capture control LV x
Minimum
amplitude RV
0.7 V 0.7 V x x x
Minimum
amplitude LV
x
Threshold test start 2.4 ... (0.6) ... 4.8 V 3.0 V x x x
RV safety margin 0.3 ... (0.1) ... 1.2 V 0.5 V x x
LV safety margin 1.0; 1.2 V 1.0 V x
Search type Interval; time of day Time of day x x x
Interval 0.1; 0.3; 1; 3; 6; 12; 24 h 24 h x x x
Time of day 00:00 ... (00:10)
... 23:50 hh:mm
00:30 hh:mm x x x
Parameter Range of values Standard SR DR HF
Atrial overdrive
pacing
OFF; ON
With ON: maximum over-
pacing rate 120 bpm, mean
rate increase approximately
8 bpm, rate decrease after
20 cycles
OFF x x