User Manual

Table Of Contents
35
Parameters
Preset Programs
Refractory period A AUTO AUTO x x
Refractory period RV 250 ms 250 ms 300 ms x x x
Refractory period LV 200 ms 200 ms 200 ms x
Mode switching ON ON x x
Onset criterion 5-out-of 8 5-out-of 8 x x
Resolution criterion 5-out-of 8 5-out-of 8 x x
Intervention rate 160 bpm 160 bpm x x
Switches to DDIR DDIR x x
The basic rate with mode
switching
+10 bpm +10 bpm x x
Rate stabilization with
mode switching
OFF OFF x x
PVARP AUTO (Start
250 ms)
225 ms) x x
PVARP after PVC 400 ms Automati-
cally set
—xx
Capture control A ON ON OFF x x x
Capture control RV ON ON OFF x x
Capture control LV ON ON OFF x
Atrial overdrive pacing OFF OFF x x
Vp suppression OFF OFF x
IEGM recording (HAR) ON AT OFF x x x
IEGM recording (HVR) ON ON OFF x x x
Home Monitoring OFF OFF OFF x x x
Parameter Factory
setting
Standard Safe
program
SR DR HF