User Manual
Table Of Contents
- Enitra 6/8
- Table of Contents
- 1 Product Description
- 2 General Safety Instructions
- 3 Implantation
- Implantation Procedure
- Having parts ready
- Keeping an external defibrillator ready
- Unpacking the device
- Checking parts
- Implantation site
- Overview: Implanting
- Avoid damage to the header
- Preventing short circuits in the header
- Keeping distance between leads
- Connecting the lead connector to the device
- Applying the programming head
- Establishing wandless telemetry
- Auto-initialization
- Behavior during auto- initialization
- Precautionary Measures while Programming
- Checking the device system
- Performing standard tests and monitoring the patient
- Do not interrupt wandless telemetry during a treatment
- Cancelling telemetry
- Avoiding critical parameter settings
- Manually setting lead polarity
- Setting sensing
- Setting the sensitivity
- Preventing device-induced complications
- Preventing conduction of atrial tachycardia
- Phrenic nerve stimulation that cannot be terminated
- Avoiding risks in the case of exclusive left ventricular pacing
- If an ICD is implanted at the same time, do not permit unipolar pacing
- Recognizing lead failure
- Consider power consumption and service time
- Magnet Response
- Follow-up
- Patient Information
- Replacement Indications
- Explantation and Device Replacement
- Implantation Procedure
- 4 Parameters
- 5 Technical Data
30
Parameters
Pacing and Sensing
Pacing and Sensing
Pulse amplitude and
pulse width
Sensitivity
Atrial capture control
Ventricular capture control
Atrial overdrive pacing
Parameter Range of values Standard SR DR HF
Pulse
amplitude A/RV/LV
0.2 ... (0.2) ... 6.0 ... (0.5)
... 7.5 V
3.0 V x x x
Pulse width A/RV/LV 0.1 ...(0.1) ... 0.5 ... (0.25)
... 1.5 ms
0.4 ms x x x
Parameter Range of values Standard SR DR HF
Sensitivity AUTO; 0.5 ... (0.5) ... 7.5 mV AUTO x
Sensitivity A AUTO; 0.1 ... (0.1) ... 1.5 ...
(0.5) ... 7.5 mV
AUTO x x
RV sensitivity AUTO; 0.5 ... (0.5) ... 7.5 mV AUTO x x x
LV sensitivity OFF; AUTO; 0.5 ... (0.5) ...
7.5 mV
AUTO x
Parameter Range of values Standard SR DR HF
Atrial capture control ATM (monitoring only); ON;
OFF
ON x x
Minimum amplitude 0.5 ... (0.1) ... 4.8 V 1.0 V x x
Threshold test start 2.4 ... (0.6) ... 4.8 V 3.0 V x x
Safety margin 0.5 ... (0.1) ... 1.2 V 1.0 V x x
Search type Interval; time of day Time of day x x
Interval 0.1; 0.3; 1; 3; 6; 12; 24 h 24 h x x
Time of day 00:00 ... (00:10) ...
23:50 hh:mm
00:30 hh:mm x x
Parameter Range of values Standard SR DR HF
Capture control RV ATM (monitoring only); ON;
OFF
ON x x x
Capture control LV x
Minimum
amplitude RV
0.7 V 0.7 V x x x
Minimum
amplitude LV
x
Threshold test start 2.4 ... (0.6) ... 4.8 V 3.0 V x x x
RV safety margin 0.3 ... (0.1) ... 1.2 V 0.5 V x x
LV safety margin 1.0; 1.2 V 1.0 V x
Search type Interval; time of day Time of day x x x
Interval 0.1; 0.3; 1; 3; 6; 12; 24 h 24 h x x x
Time of day 00:00 ... (00:10)
... 23:50 hh:mm
00:30 hh:mm x x x
Parameter Range of values Standard SR DR HF
Atrial overdrive
pacing
OFF; ON
With ON: maximum over-
pacing rate 120 bpm, mean
rate increase approximately
8 bpm, rate decrease after
20 cycles
OFF x x