User Manual

Table Of Contents
40
Technical Data
Battery Data
Mean service times D, DR
For dual-chamber devices of the 4 series the following times result when set to
DDD(R), with a basic rate of 60 bpm and a pulse width of 0.4 ms at an impedance of
500 Ω:
For dual-chamber devices of the 8 series the following times result when set to
DDD(R), with a basic rate of 60 bpm and a pulse width of 0.4 ms at an impedance of
500 Ω:
Mean service times HF
For triple-chamber devices of the 8 series the following times result when set to
DDDR with a basic rate of 60 bpm, 100% biventricular pacing and a pulse width of
0.4 ms at an impedance of 500 Ω:
Amplitude Pacing Average service time
A: 2.5 V
RV: 2.5 V
100% 9 years, 10 months
50% 12 years, 4 months
A: 3.0 V
RV: 3.0 V
100% 7 years, 11 months
50% 10 years, 8 months
A: 5.0 V
RV: 5.0 V
100% 8 years, 1 months
Amplitude Pacing Average service time
A: 2.5 V
RV: 2.5 V
100% 9 years, 4 months
50% 11 years, 4 months
A: 3.0 V
RV: 3.0 V
100% 7 years, 8 months
50% 10 years
A: 5.0 V
RV: 5.0 V
100% 3 years, 2 months
Amplitude Pacing Average service time
A: 2.5 V 10% 9 years, 8 months
RV: 2.5 V
LV: 2.5 V
100%
A: 3.0 V 10 % 8 years
RV: 3.0 V
LV: 3.0 V
100%
A: 5.0 V
RV: 5.0 V
LV: 5.0 V
100% 2 years, 6 months