Apply the programming head on the patient over the
device.
Select: MoreLumax TelemetryPGH.
Power Consumption Consideration:
SafeSync telemetry requires somewhat more power than
telemetry via the programming head. Power consumption during
implantation corresponds to approximately 10 days of service
time and consumption during 20-minute follow-up corresponds to
approximately 3 days. As a result:
Do not establish SafeSync Telemetry sessions
unnecessarily.
After 5 minutes without input, SafeSync Telemetry
switches to the economy mode. In order to re-establish
telemetry from the economy mode, select “Close” in the
pop-up window with the patient’s name:
Check the battery capacity of the device at regular
intervals.