Lumax Technical Manual 71
The SafeSync symbol is displayed in the navigator and
the signal strength is displayed in green bars on the
information line. The SafeSync telemetry status is shown
in five increments: 1 bar = 20% up to 5 bars = 100%. A
weak contact shows only one bar in green, whereas all
five bars are shown in green when there is optimal
contact. The bars are shown in gray if RF telemetry is
lost. The display should have at least three green bars.
Otherwise, it is recommended to reposition the
programmer until an adequate signal strength is
achieved.
Remove the programming head.
To restore the SafeSync Telemetry after an interruption
during follow-up or after a programmer restart: select
SpecialContinue RF session on the start screen to
restore the most recent active session. This SafeSync
Telemetry session must not have been inactive for longer
than five minutes, since the device's SafeSync Telemetry
activity is automatically switched off after five minutes of
inactivity.
Note: If the SafeSync Telemetry session has been
inactive for more than 5 minutes, the programming
head must be reapplied to reinitialize the SafeSync
Telemetry.
To End SafeSync Telemetry contact:
To end the SafeSync Telemetry session:
End Follow-up session
To switch to telemetry via the programming head (PGH):