70 Lumax Technical Manual
2. Device Features
The Lumax family feature set is presented under the following
sub-headings: Tachyarrhythmia Detection, Tachyarrhythmia
Redetection/Acceleration, Tachyarrhythmia Therapy,
Tachyarrhythmia Termination, Bradycardia Therapy, EP Test
Functions and Special Features. The features apply to all
members of the Lumax family except where specifically
referenced differently.
CAUTION
Programmed Parameters – Program the device
parameters to appropriate values based on the patient’s
specific arrhythmias and condition.
2.1 SafeSync Telemetry
The Lumax 700/740 and 600/640 models offer ”wandless”
communication between the device and the programmer by using
radio frequency (RF) telemetry, in addition to the currently
available telemetry used by applying the programming head
(PGH) over the implanted device. This function is called SafeSync
Telemetry.
SafeSync Telemetry can be used with the Renamic programmer
or with the ICS 3000 programmer (using the SafeSync Module,
an external communication module).
To Establish SafeSync Telemetry contact:
The programmer (or the SafeSync module) must be no more than
9 feet (3 m) from the device; ideally there should be no obstacles
between the patient and the programmer.
Switch on RF telemetry on the programmer. Select:
Preferences→ConnectivityRF telemetry
InterrogationON. Or, select: MoreLumax
TelemetryRF, during the follow-up.
Apply the programming head for about 2 seconds until
successful initialization is displayed on the programmer: