User Manual

Table Of Contents
Lumax Technical Manual 65
Description Results
Complication Rate 11.1% (1/9)
Appropriate Atrial Sensing Rate
1
97.6% (41/42)
Appropriate Atrial Sensing during 24-hour
Holter Test
100% (9/9)
Appropriate Atrial Sensing during
Exercise Treadmill Test
100% (6/6)
Detection and Conversion of Ventricular
Tachyarrhythmias
2
100% (68/68)
1.6.5.2 European Clinical Study
1.6.5.2.1 Patients Studied
The European Deikos A+/Kainox VDD lead clinical study involved
82 patients (66 males and 16 female) with a mean age of 61.8
years (range: 29 to 84 years).
42.7% presented with
monomorphic ventricular tachycardia as their primary
tachycardia.
1
The investigator determined the appropriateness of atrial sensing. The rate is
determined by the number of appropriate atrial sensing evaluations divided by
the total number of evaluations.
2
Conversion data were collected in the clinical study for both induced and
spontaneous ventricular tachyarrhythmia episodes. Therefore, both types of
tachyarrhythmia episodes were included in the analysis.