User Manual

Table Of Contents
60 Lumax Technical Manual
Figure 5: Safety Event Rate Kaplan Meier
Analysis
The safety event rate for a 12-month duration was 14.8% for
Group 1 (HM) and 16.4% for Group 2 (Control), with a non-
inferiority p-value of 0.005. Therefore, the safety event rate for
HM Group was non-inferior to the safety event rate for the Control
Group within 5%. The upper, one-sided 95% confidence bound
for the difference was 2.7%.
A rejection of the null hypothesis indicates that the safety event
rate for Group 1 (HM) is equivalent (non-inferior) to that of Group
2 (Control).
1.6.4.3.3 Secondary Endpoint 1: Early Detection of Cardiac
Ev
ents (AF, VT & VF)
The purpose of secondary endpoint 1 was to compare AF, VT
and VF events between Group 1 and Group 2 in terms of the
number, categories, and detection time relative to onset.