Anomaly Possible Effect on
Patient or Implant
Procedure
Lumax Application (Programmer
Software)
Impact on
Lumax 700/740 &
600/640
An ineffective
pacing program
may continue
during the
threshold test
at
programming
distances close
to the loss of
communication
. Lifting the
wand corrects
the problem.
ID# 17903
Limited potential effect
on patient, non-capture
is quickly recognized
and removal of
programming wand is
typical response. This
anomaly can only occur
during manual threshold
testing (in physician’s
office).
N/A