204 Lumax Technical Manual
Anomaly Possible Effect on
Patient or Implant
Procedure
Lumax Application (Programmer
Software)
Impact on
Lumax 700/740 &
600/640
For long patient
names
involving
umlauts the
transmission of
patient data
may suppress
the patient
name (empty
field).
ID# 393925
No effect on patient.
Functional/Usability
Issue (missing,
insufficient user
guidance, unexpected
behavior):
Unexpected/Unspecified
behavior without clinical
relevance or negative
therapeutical/diagnostic
implication. Acceptable
residual risk resulting
from low severity and
low probability of
occurrence.
N/A
Dual Chamber
devices only:
When
programmed to
DDDR mode
and with
sensor activity
it is possible for
a sudden rate
drop at start of
a capacitor
reformation
procedure.
After the
reformation the
pacing rate
again sensor
indicated rate.
ID# 19607
Limited effect on patient,
temporary rate drop to
programmed post-shock
pacing rate ensures
sufficient support.
N/A