202 Lumax Technical Manual
Anomaly Possible Effect on
Patient or Implant
Procedure
Lumax Application (Programmer
Software)
Impact on
Lumax 700/740 &
600/640
The
transmission
date for
Periodic IEGM
messages for
Remote
Follow-Up
(Home
Monitoring)
may be off by
one day.
ID# 412325
No effect on patient.
Functional/Usability
Issue (missing,
insufficient user
guidance, unexpected
behavior):
Unexpected/Unspecified
behavior without clinical
relevance or negative
therapeutical/diagnostic
implication. No message
is lost, no data is
altered. Accepted due to
low probability of
occurrence and very low
severity
N/A
Event Counter
Statistics may
be slightly off
by single
counts of Vx
events
dependent on
internal
statistics state
transitions.
This is a rare
event.
ID# 410925
Limited effect on patient.
Diagnostic Data Issue
(Interrogation, Display,
Printout):
Display/printout of
incorrect diagnostic
information without risk
of incorrect diagnosis.
Acceptable residual risk
resulting from low
severity and low
probability of
occurrence.
Product safety and
performance are
only marginally
affected.